WSI-YP2000 LCD display Infant Incubator with X-ray cassette
Product Introduction
Two control modes: Air mode and baby mode controlled by micro-computer;
The information can be stored, consulted and indicated by the LCD display screen;
Humidity and oxygen concentration display and servo control;
>37℃temperature set function;
Double wall hood, air circulation system in the way of air outlet from front and back;
X-ray cassette under the bassinet;
Embedded integrated sensor module, integrated water reservoir in the shape of drawer;
Inclination of bassinet is adjustable;
The height of whole unit can be adjusted;
Failure alarm indication;
RS-232 connector;
DC motor.
Basic Configuration
Baby compartment, main body, vertical height adjustment stand (VHA stand), the controller, I.V. Pole and shelf.
Main Parameters
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 1300VA
Mode of control: Air mode and baby mode controlled by micro-computer
Air temperature control range: 25℃-37℃(override mode 37-39℃)
Baby temperature control range: 34-37℃(override mode 37-38℃)
Accuracy of skin temperature sensor: ±0.2℃
The temperature difference between incubator temperature and average incubator temperature(under steady temperature condition): ≤0.5℃
Temperature uniformity: ≤0.8℃(mattress is on the horizontal position)
Inclination of the bassinet: ±8°
Noise inside hood:≤45dB(A)
Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, Motor Fan Failed Alarm, Power failure Alarm, sensor box position alarm, tank position alarm, system alarm and so on.
Humidity display range: 0%RH~99%RH
Humidity control range: 0%RH~90%RH
Accuracy of humidity display: ±5%RH(environment temperature is 25℃, environment humidity is 45%RH±5%RH)
Indication range of oxygen concentration: 0%~99%
Control range of oxygen concentration: 20%~60%
Accuracy of oxygen display concentration: ±2%(oxygen concentration sets ≤25%)
±3%(oxygen concentration sets >25%)
Accuracy of oxygen control concentration: ±4%
Range of weigh: 100-8000g(weighing system)
Resolution for weight indication: 1g(weighing system)
Accuracy of weigh: ±1%(weighing system)
Environmental temperature:
Operating range: 20 ℃~30℃
(Set temperature should be higher 3℃ than the environment temperature)
Ambient air movement rate: <0.3m/s
Product Packaging:
Package: each unit is packed into two cartons
Dimension (Main body): 1110mm x 710mm x 888mm
(Cabinet): 1029mm x 697mm x 826mm
Gross Weight (Main body): 70Kg
(Cabinet): 80Kg