Function: Configured with single table and single tube, radiography, satisfying ordinary clinic usage
Standard configuration: 1 controller, 1 generator, 1 tube, 1 all-in-one radiographic table and 4 high-voltage cables Specifications: Power: A.C. 380V±10%,frequency 50HZ±1HZ, Capacity≥50KW; Power resistance:≤0.30Ω,Earth resistance≤4Ω
High Voltage Generator: Fluoroscopy: 0.3-5mA , step enters:0.1mA, 50-110KV, step enters: 1 KV, continuous exposure time ≥3h Radiography: 30-500mA divided adjustable, 45-125KV, step enters: 1 KV,0.01-6.3s, totally 26 steps are adjustable
X-Ray Tube Assembly: Diagnostic table: Small focus: 1×1, 20KW, Large focus:2×2, 40KW, normal revolving speed 2800r/min, heat capacity 140KHU Radiographic table: Small focus: 1×1, 30KW, Large focus:2×2, 50KW, normal revolving speed 2800r/min, heat capacity 140KHU
Collimator: Photographic table: Manual and leafy type
Multi-function Radiographic Table: Movement of the tabletop: Longitudinal ±450mm;transverse: ±120mm The tabletop height: 700mm Moving grid: Grid density: 28L/cm, Grid ration: 8:1, Focusing distance: 100mm, moving range: ≥600mm Film’s specifications:5″×7″,8″×10″,10″×12″,11″×14″,12″×15″,14″×14″,14″×17″ Movement of X-Ray tube assembly: To go around the horizontal axle ±180°,up and down: 1100mm